Decontamination and reprocessing of units and accessories

All Neurotarc Products has very similar accessories, below is the common approach for cleaning after single use.



Main Unit (the electical device)




  • Suitable for multiple users. Should be cleaned thoroghly when passed to another person. Advised to be wiped after each use.

  • Cleaning Instructions: We recommend using Clinell Universal Wipes or WW1504 Patient Cleansing wet Wipes, suitable for hands and equipment wiping.

  • Cleaning Warnings: Do not use cleaning sprays or alcohol based cleaning solutions. Do not use UV or steam – based sterilising equipment! 

    Alcohol-based solutions should not cause any functional damage to the unit, if applied gently and not over exposed. There is a high possibility of label and surface deterioration of the unit if alcohol-based solutions are used. Never immerse the unit in any fluid! Use wipes only. If you cannot avoid alcohol-based solutions, never spill alcohol-based solution over the unit areas, the alcohol wipes should never contain more than 70% of spirit, wipe off quickly, and do not over expose.



Lead wires (dual or single conductor)




  • For single patient use only! Advised to be cleaned thoroghly after each use.

  • Cleaning Instructions: We recommend using Clinell Universal Wipes or WW1504 Patient Cleansing wet Wipes, suitable for hands and equipment wiping.

  • Cleaning Warnings: Do not use cleaning sprays or alcohol based cleaning solutions. Do not use UV or steam – based sterilising equipment! 

    Alcohol-based solutions should not cause any functional damage to the wire, if applied gently and not over exposed.  Never immerse the wire in any fluid! Use wipes only. The alcohol wipes should never contain more than 70% of spirit, wipe off quickly, and do not over expose.



Probes (vaginal or rectal)




  • For single patient use only! Should be cleaned thoroghly after each use.

  • Cleaning Instructions: Clean the VeriProbe before the first use. Clean the VeriProbe before and after use.  Wash the probe gently in mild soapy water, rinse and make sure the probe is completely dry before returning to storage in the plastic bag. For the detailed instructions, follow the instructions for use supplied with the probe.



Skin electrode pads




  • For single patient use only! Should be cleaned thoroghly after each use.

  • Cleaning Instructions: don't clean, avoid contamination.  Do not use alcohol wipes, gels or other liquids to clean any side of the electrodes. 2) Do not apply to broken skin. 3) Stop using if irritation occurs!

    .