Verity Medical Logo

Detailed specifications

 

     
     
Products  

NeuroTrac™ Labour TENS

Click here to return to the home pageClick here to view our TENS, Stimulators, Probes and EMG unitsClick here to view our range of Skin Electrodes and information on types and sizes of Skin ElectrodesClick here to view EMG, TENS, STIM and product FAQ'sClick here to view detailed information on TENS, Stimulation (STIM or EMS) and EMGClick here to view a list of Protocols developed by trained clinicians around the worldClick here to view retail and distributor sales, OEM Branding and custom product designClick here to download photos, brochures, manuals and protocols

 

 

NeuroTrac Labour TENS Unit

 

Remote

 

 

 

 

Specifications

TENS
1. Dual channel: individually isolated circuits.
2. Amplitude: 0-80 mA; indication only; actual mA will tend to be less than indicated due to electrode impedance.
3. Type: Constant Current.
4. Waveform: Asymmetrical, rectangular bi-phasic with zero DC current.
5. Pulse width: 220mS [2% accuracy].
6. Pulse Rate: 90Hz (Continuous mode) [2% accuracy].
7. Mode: Continuous (Boost mode).
8. Burst mode: Bursts of 9 pulses [200 mS] at 150 Hz, repeating twice every second (Rest mode).
9. Ramp up Time: 0.8 seconds from Rest to Boost
10. Time duration: continuous until switched off.
11. Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2 volts the battery symbol will flash on/off once every second.
12. If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not turn on.
13. Open Electrode Detect: If an open circuit is detected at the output of channel A or B the output current will be reset at zero.
14. Physical dimensions: 108 x 62 x 23 mm.
15. Weight: 0.09KG without battery, 0.14KG with battery.
16. Environmental conditions for storage & transport:
        -10 to +50 degrees Centigrade
        0-90% Humidity.